Infinity DB Series Blackbody Overview:

Santa Barbara Infrared's Infinity Series of Differential Blackbodies (DB Series) are designed to support the most critical IR testing requirements. The Infinity blackbodies offer the best performance available. Fast, stable, uniform, accurate and reliable, these blackbody systems provide performance enhancements and features that set them apart from previously available systems. A key feature is the new iProbe from SBIR. This intelligent temperature sensor is calibrated independently of the blackbody system. To re-calibrate the system you need only exchange the probe with a recently calibrated one. No special equipment is required thus minimizing downtime. Our blackbodies are designed for use as flood sources for non-uniformity correction or sensor calibration, and as differential sources for MRT, MTF, SiTF and other standard IR sensor characterization tests.








Features:
  • i-probeIntelli-Probe (iProbe)
    The iProbe from SBIR is an intelligent temperature sensor that is calibrated independently of the blackbody system. To re-calibrate the system you need only exchange the probe with a recently calibrated one. No special equipment is required thus minimizing downtime.

  • Accuracy

    Ultra-stable, highly responsive iProbes are used to provide very precise control and accurate system performance. The iProbes are calibrated to an accuracy of under 0.010ºC. System uncertainty including calibration and stability is +/- 0.01ºC or +/- 0.1% of set delta T whichever is greater.

  • Uniformity
    The Infinity blackbody incorporates design and construction features that maximize spatial uniformity and minimize convection influences resulting in typical surface uniformity of > 98% of set differential temperature or 0.01ºC, whichever is greater over 90% of the emitting surface area.

  • Emissivity

    A surface treatment is used to create a spectrally flat emissivity of greater than .970 for 2µ to 14µ.

  • Milli-Kelvin Temperature Stability

    SBIR has developed unique temperature measurement and analog-to-digital conversion circuitry with integrated control logic that results in extremely responsive temperature control. True set point and accurate milli-kelvin control have been achieved allowing for precision testing at contrast levels of 1 milli-kelvin.

  • Multi-Function Controller

    The controller features a touch screen display and supports remote computer control over Ethernet or RS-232. The controllers are also available with an optional IEEE 488 GPIB Interface.

    The controller will also operate all of the SBIR peripherals: target wheels, visible illuminators, pulsed laser diodes, range slides, source slides and filter wheels.
    Controller

  • Automatic Radiometric Compensation

    Attenuation caused by optics, differences in emissivity, ambient temperature change, and waveband of interest result in system output that is not radiometrically correct. When operating the controller in the radiometric mode, compensation for these factors is automatically applied, adjusting the blackbody as required to maintain the proper system output.

  • IRWindows™ Advanced EO Test and Analysis Tool

    The Infinity Series is optimized for seamless integration with the SBIR IRWindows™ test system. IRWindows™ is an advanced software tool that automates the setup, execution, data collection and results analysis for industry standard performance testing of infrared, visible and laser sensor systems.

  • Configurations

    The Infinity Differential Blackbody system includes a temperature controller (19” rack mount or desktop configuration), blackbody head, interconnect cable, power cable and manual. The blackbody heads are available with emitting surface sizes including 4”, 6”, 8” and 12” square. Standard and optional temperature ranges can be viewed in the Specifications section.
Infinity DB Series BlackbodyOptions:

  • IEEE Std. 488 Interface (GPIB)
    The thorough implementation of the ANSI/IEEE Std. 488.1-1987 allows for computer control of all blackbody, calibration, illuminator, and target wheel functions.

  • Radiometric Differential Temperature
    Differential temperature errors due to optical attenuation or emissivity differences can be corrected with this feature.

  • Illuminator
    A visible alignment light moves into and out of the optical path to backlight the target. The illuminator position is controlled from the front panel or via the computer interface options. Light level luminance is preconfigured and adjustable at 10 various intensity set points. This feature is very useful for visible alignment of the system. IR-to-visible boresight tests can be made with uncompromised accuracy because the target is stationary during IR to visible transition.

  • Targets
    Many standard and custom infrared targets are available. These targets can be supplied with either emissive or reflective coatings. See the Target section for details.

  • Target Wheels
    Multiple targets can be rapidly and precisely positioned in front of the blackbody. Target position can be set from the front panel of the controller or via the computer interface. For details on the full line of target wheels see SBIR’s Target Wheel section.

  • Rack Mounting
    The blackbody controller is designed for bench-top or standard 19 inch rack use. Rack mounting ears come standard with the instrument. A rack slide kit is also available.

Specifications:

Available Sizes 4”, 6”, 8”, 12”
   
Standard Temperature Range A      -25ºC to +75ºC ΔT, 0ºC to 100ºC Absolute
Optional Range B -25ºC to +150ºC ΔT, 0ºC to 175ºC Absolute
Optional Range C1 -20ºC to +75ºC ΔT, -40ºC to +100 Absolute
Optional Range D1 -20ºC to +150ºC ΔT, -40ºC to +175 Absolute
Optional Range E2 -25ºC to +55ºC ΔT, 0ºC to +80ºC Absolute
   
Emissivity > 0.970, 2 µ to 14 µ
   
Uniformity

98% of set ΔT or 0.01ºC, whichever is greater over 90% of emitting surface area

   
Absolute Accuracy ±0.007ºC for 0ºC<T<50ºC or ±0.0005 (T-25) elsewhere
Differential Acccuracy ±√[(T1 error)2+(T2 error)2] where error is ±0.007ºC
for 0ºC<T<50ºC and ±(T-25)(0.0005) elsewhere
   
Stability

σT≤0.001ºC (0ºC to 50ºC), σT≤0.002ºC (-40ºC to 0ºC and 50ºC to 100ºC),
σT≤0.003ºC (110ºC to 175ºC)

   
Display Resolution 0.001ºC or 0.0001ºC
Setpoint Resolution 0.001ºC
Ready Indicator Selectable ±0.001ºC to ±5.000ºC
   
SLEW RATE3  
   
Approximate Heating Rate +0.40ºC/sec. at 25ºC
Approximate Cooling Rate -0.20ºC/sec. at 25ºC
Settling Time <45 seconds
   
ENVIRONMENTAL  
   
Operating Temperature 0˚C to 50˚C
Storage Temperature -20˚C to 70˚C
Relative Humidity 5% to 95%, non-condensing
Power Requirements 100-120/220-240 VAC at 60/50Hz
Maximum Power Consumption 850W-1600W depending on size and temperature range
Approximate Blackbody Weight DB-04—15 lbs., DB-06—25.0 lbs., DB-08—40.0 lbs., DB-12—50 lbs.
Approximate Controller Weight 15.0 lbs.

Notes:

  1. For use in a thermal chamber from (-20˚C thru +50˚C ambient)
  2. Specific only to 12” model
  3. Slew rates vary depending on blackbody size and environmental conditions

* Specifications subject to change without prior notice

PDF Application Drawings





Infinity DB Series BlackbodyPDF Data Sheet:

See PDF Datasheet
Infinity DB Series BlackbodyOrder Information:

SBIR classifies blackbody model numbers based on six different features that a customer can choose from. Features include blackbody size, temperature range, country’s power cable type, specified cable length, communications interface options and configuration options.

For example, a DB Series blackbody with a temperature range of 0ºC to +100ºC, 115/60 Hz US country’s input voltage, 10 foot cable length, IEEE 488 GPIB communication interface and a target support plate for mounting stand-alone targets has the following part number: DB-04-A-L-10-EG-P

DB Order Info

Differential Blackbody: DB

Differential Blackbody Size: 04", 06", 08" or 12"

Temperature Range:

  • A = -25ºC to 75ºC ΔT, 0ºC to 100ºC Absolute
  • B = -25ºC to 150ºC ΔT, 0ºC to 175ºC Absolute
  • C1 = -20ºC to 75ºC ΔT, -40ºC to 100ºC Absolute
  • D1= -20ºC to 150ºC ΔT, -40ºC to 175ºC Absolute
  • E2= -25ºC to 55ºC ΔT, 0ºC to 80ºC Absolute

Notes:

  1. For use in a thermal chamber from (-20ºC thru +50ºC ambient)
  2. Specific only to 12” model

Power Cable :

  • H- Europe (Germany, Netherlands, France, Sweden)
  • I- Australia & China
  • J- UK & Ireland
  • K- Italy
  • L- US & Japan

Cable Length: 10 ft., 25 ft., 50 ft.

Interface Options:

  • ES- Ethernet and RS-232
  • EG- Ethernet and IEEE-488 GPIB

Target Probe Options:

  • P- Target Support Plate
  • Q- No Target Support Plate
  • R- Radiometric Compensation
Please contact the SBIR sales team at (805) 965-3669 to ensure the proper part number and to receive a quotation.

Information subject to change without notice