FEATURES
| OPTIONS | DIMENSIONS
SPECIFICATIONS | ORDER INFO
Santa
Barbara Infrared's cavity blackbody is built around a new and innovative
cavity construction that significantly improves the uniformity and slew
rate of the cavity, improving test accuracy and reducing test times.
Additionally,
this new product incorporates the advantages of SBIR's breakthrough Smart
Blackbody architecture. The design of the Smart Blackbody creates
a system with inherently more stable and accurate temperature control.
Additionally, foolproof blackbody interchangeability, high immunity to
EMI, ease of calibration, and non-critical cabling all make the Smart
Blackbody an easy-to-integrate component into a large test system.
These
advanced systems are designed for ease of use and maximum reliability.
SBIR's Model 4100 blackbody is the ideal choice for high temperature applications
where high performance is a requirement.
F
E A T U R E S
Superior
Uniformity and Speed
The unique cavity construction of the Model 4100 preserves the high emissivity
of traditional cavity blackbody designs, but greatly improves the uniformity
and slew rate of the system. Cooling rates are markedly faster, significantly
reducing test times.
Smart
Blackbody Architecture
SBIR's Smart Blackbodies put the temperature measurement and calibration
electronics in the blackbody head rather than in the control electronics
package. With this new architecture, temperature measurement information
from the blackbody to the controller is serial digital data rather than
an analog temperature sensor signal. The result is significantly improved
performance: lower noise operation, more stable temperature control, improved
immunity to EMI, and an easier integration task for the test equipment
engineer.
The
calibration of the blackbody head is independent of the controller, so
heads and controllers can be swapped without recalibration.
High
Temperature Range
The cavity temperature range extends from 50°C to 1000°C.
High
Accuracy and Stability
The microprocessor-based temperature controller is designed for optimum
accuracy and stability. A high resolution enhanced PID temperature servo
provides zero drift and low jitter control of the emitting surface's temperature.
Ease
of Calibration
Calibration is fast and easy. Temperature corrections are made "closed
case", without internal trims, adjustments, or reprogramming of EPROMs.
All adjustments can be made directly from the front panel or through the
remote interface. Calibration points are independent of each other, allowing
the user to touch up a single calibration point.
Built
In Test (BIT)
An extensive BIT assures that the blackbody is functioning properly during
use. BIT is automatically invoked on power up, and a continuous background
(transparent to the user) BIT is performed during operation.
O
P T I O N S
IEEE
Std. 488 Interface (GPIB)
The thorough implementation of the ANSI/IEEE Std. 488.1-1987 allows computer
control of all blackbody functions, calibration, and optional features
such as target wheels and illuminators. (Option G)
RS-232
Serial Interface
An optional serial interface provides computer control of all blackbody,
calibration, illuminator and target wheel functions. (Option S)
Aperture and Filter Wheels
A manually-operated aperture wheel is included with the blackbody. Motorized
wheels are available to enable multiple apertures or filters to be rapidly
and precisely positioned in front of the blackbody. Selection can be made
from the front panel of the controller or via the optional remote interface
(IEEE 488 or RS-232).
Modulator
A mechanical chopper can be used to modulate the blackbody radiation.
Dynamic range of the chopper is 1000:1. A variety of blade sizes and speeds
are available. The modulator can be controlled from the front panel or
through the remote interface.
Rack Mounting
The blackbody controller is designed for bench-top or standard 19 inch
rack use. Rack mount ears are included with the instrument. A rack slide
kit is also available.
Handheld Remote Panel
An optional remote control is available. All instrument front panel functions
can be accessed using this handheld unit. It is connected to the temperature
controller with a 10 foot cable.
D
I M E N S I O N S
S
P E C I F I C A T I O N S
Preliminary
Specifications:
| Temperature
range |
|
50°C
to 1000°C |
| Exit
aperture |
|
1"
diameter |
| Setpoint
resolution |
|
0.1°C |
| Display
resolution |
|
0.1°C |
| Slew
rate, including settling time |
|
100-1000°C,
<60 minutes
1000-100°C, <85 minutes |
| Stability
|
|
±0.1°C
short term
±0.5°C long term |
| Emissivity
|
|
>.99 |
| Controller
Dimensions |
|
17"W
x 3.5"H x 14" D (19" rack mountable) |
| Cable
Length |
|
10'
standard, other lengths available |
| Line
Voltage |
|
Selectable
120/220 V, 50/60 Hz |
| Detailed
Outline Drawings in PDF Format |
 |
 |
 |
4100
Series
Controller |
4100
BB Head |
Electromechanics
Controller |
If you
don't yet have the free Adobe Acrobat Reader, get it here:

O
R D E R I N F O R M A T I O N
Options
to the Model 4100 Cavity Blackbody are specified as suffixes to the model
number. IEEE-488 interface is Option G. RS-232 interface is Option S.
Variable Frequency Chopper is Option C, a Motorized Aerture Wheel is Option
P, a Motorized Filter Wheel is Option F, and a Shutter is Option U. Other
features such as non-standard cable length or rack slide kits should be
specified. Contact SBIR for assistance
in selecting standard or custom systems.
Information